    Ultra Micro-volume Spectrophotometer TL6
    TL6 Ultra Micro-volume Spectrophotometer
    TL6 is the latest Ultra Micro-volume Spectrophotometer of PERSEE ANALYTICS. It is a small portable instrument for fast analysis of nucleic acid and protein concentration by pipette sample droplet. You can drop the sample (sample volume= 0.5uL~2uL) directly onto the measuring head for sample analysis. When the analysis is complete, please use pipette to recycle the sample or the air-laid paper remove the sample (no need to use the cuvettes or capillary). All the steps are simple and fast!
    Description:
    TL6 is the latest Ultra Micro-volume Spectrophotometer of PERSEE ANALYTICS. It is a small portable instrument for fast analysis of nucleic acid and protein concentration by pipette sample droplet. You can drop the sample (sample volume= 0.5uL~2uL) directly onto the measuring  head for sample analysis. When the analysis is complete, please use pipette to recycle the sample or the air-laid paper remove the sample (no need to use the cuvettes or capillary). All the steps are simple and fast!
    The TL6 is equipped with built-in methods for nucleic acid and protein quantification analysis, including UV adsorption method, Bradford dye binding assay and the color method (Lowry, BCA and Biuret). In addition, the continuous wavelength design allows the users to customize their own detection methods and store them in the instrument. TL6 equipped with a large blue Backlit LCD display and operation panel, can help the users to set the instrument and carry on all the detection process directly without an additional computer. This design is helpful for saving laboratory space. TL6 comes with the calibration function and can be used for sample analysis anytime and anywhere without preheating.
    ​Features & Functions:
    ​Patented Ultra Micro-volume sample detection technology
    High performance Ultra micro-volume cell technology
    Dual-channel CCD signal detection technology
    Unique functional module design
    Brand new hydrophobic sample test surface
    You only need micro sample volumes
    Fast test speed
    Easy to operate
    Flexible test method
    Longer life optical components
    Multiple data export methods
    Applications:
    The TL6 Ultra Micro-volume Spectrophotometer is an ideal instrument for Nucleic acid, protein quantification and monitoring the growth of bacterial cultures analysis in public health, life sciences and research industries. It has become a conventional instrument for the modern molecular biology laboratory. Meantime, it also can fully achieve spectral scanning and quantitative measurement functions as a standard UV  spectrophotometer with small dimension and lower power consumption.
    ​Disease control
    University institutes
    Agricultural inspection agency
    Exit Inspection and Quarantine
    Life sciences-related industries
    Specifications:
    Key performance specifications
    Optical path
    0.4 mm, 1 mm ; Switch Manually
    Optical system
    →Click here to know more about
    Optical Path System←
    Double beam
    Light Source
    Long-life Xenon lamp
    Receiver
    Dual Array CCD, 2048 Pixels
    Wavelength range
    190 nm-1100 nm
    Work wavelength range
    200 nm-900 nm
    Wavelength accuracy
    ±1.0 nm
    Wavelength repeatability
    ≤0.3 nm
    Spectral bandwidth
    →Click here to know more about
    SBW←
    3 ± 0.6 nm
    Micro mode transmittance accuracy
    ±1 %
    Micro mode transmittance repeatability
    ≤0.5 %
    Transmittance accuracy
    ±1 %
    Transmittance repeatability
    ≤0.3 %
    Stray light
    →Click here to know more about
    Stray Light←
    220 nm (NaI): <0.5 %
    360 nm (NaNO
    2
    ): <0.5 %
    Noise
    0 % noise:≤0.1 %
    100 % noise:≤0.5 %
    Drift
    ≤0.5 %/30 min
    Baseline flatness
    ±0.005 Abs (210 nm ~890 nm,0 Abs)
    Main unit function specifications
    Micro optical path switch manually
    0.4 mm, 1 mm ; Manual switch
    Compatible 10mm cuvette
    Printer
    USB
    Network( LAN )
    LAN Connect with the LAN for network communication
    Communication Interface
    USB main interface: Support the flash memory, saving the data into the flash memory. Also can connect with the USB printer. Also can connect with the USB mouse for interface operation.
    USB slave interface: Connect with the PC for the program updating by special Persee Portable Instruments Data management software. Or download, view, export and print the data.
    Power supply(DC)
    15V   DC
    Dimension(W×H×D)
    11 × 5 × 5 in
    Weight
    2 lb