-40%

Molecular Devices SpectraMax Plus384 Microplate/Cuvette Reader/Spectrophotometer

$ 1108.8

Availability: 100 in stock
  • Custom Bundle: No
  • Modified Item: No
  • Refund will be given as: Money Back
  • Country/Region of Manufacture: United States
  • Item must be returned within: 30 Days
  • MPN: Does Not Apply
  • Return shipping will be paid by: Seller
  • Color: tan
  • All returns accepted: Returns Accepted
  • Condition: Just the Spectramax unit itself. Working fine! No issues whatsoever.
  • Model: SpectraMax Plus 384
  • Brand: Molecular Devices
  • Restocking Fee: No

    Description

    SpectraMax Plus 384 Microplate Reader A microplate spectrophotometer with a built-in cuvette port
    Introduction
    The SpectraMax® Plus 384 Microplate Reader from Molecular Devices can run both standard spectrophotometer and microplate reader applications on the same instrument. Read one sample or up to 384 in a single plate read using any standard cuvette, 12 x 75 mm test tube, or 96- or 384-well microplate. For more sample throughput, the system can be easily integrated into a full robotic system and easily adapts to fit sample processing requirements.
    Patented Pathcheck Sensor
    The SpectraMax Plus 384 reader utilizes Molecular Devices’ PathCheck® Sensor—the only patented* technology available that measures the depth (optical pathlength) of samples in a microplate. When used with SoftMax® Pro Data Acquisition and Analysis Software, it can automatically normalize the well absorbance to a cuvette equivalent pathlength of 1 cm. This is equivalent to having 96 or 384 cuvettes.
    The PathCheck sensor also allows for the elimination of standard curves. For compounds with a known extinction coefficient, concentration can be calculated directly from absorbance.
    With the PathCheck sensor, it is acceptable to have different volumes in the wells, so pipetting does not have to be accurate. Add 50–300 µL of sample or standard into each well of a 96-well plate (20 µL to 100 µL in a 384-well plate) and read. The PathCheck sensor will correct for the volume differences in all 96 or 384 wells and report 1 cm absorbance values.
    Extend the dynamic range to 6+ OD. If a well reads out of range (> 4 OD), decrease the volume in that well and re-read the plate using the PathCheck sensor. A 100 µL sample (optical pathlength ~0.3 cm) that reads 2.8 OD will be corrected to ~9.2 OD
    Pipetting errors in 96- or 384-well microplates can be detected. The PathCheck sensor measures differences in volume between the wells, so sources of error can be identified and eliminated from data, or absorbance values can be corrected.
    Quickly and easily test multi-channel dispensers and pipettors. Measuring the depth of the liquid in the well, the PathCheck sensor can determine the volume of liquid dispensed. Even 96- or 384-channel dispensers can be tested in a matter of minutes.
    Applications
    Essentially, all UV/VIS spectrophotometric assays that are read in a conventional spectrophotometer or microplate reader can be read in the SpectraMax Plus 384 reader:
    • DNA quantitation
    • Colorimetric assays
    • ELISAs/EIAs
    • Immunoassays
    • Drug dissolution profiles
    • Enzyme kinetics (e.g., Ki , Km, etc.)
    Comprehensive data analysis
    With SoftMax Pro Software, choose endpoint, discontinuous kinetics, or spectral scans; view kinetic and spectral runs in real time and collect and store data from multiple plates, cuvettes, graphs, analysis tables and notes sections in the same data file. Lab processes can be automated with the software’s interprocess messaging features and data and template information can be exported and imported to and from a LIMS system
    Technical specifications
    Photometric performance
    Wavelength range 190–1000 nm
    Wavelength selection Monochromator, tunable in 1.0 nm increments
    Wavelength bandwidth 2 nm
    Wavelength accuracy ±1.0 nm
    Wavelength repeatability ±0.2 nm
    Photometric range 0–4.000 OD
    Photometric resolution 0.001 OD
    Photometric accuracy (microplate) < ±0.006 OD ±1.0%, 0–2 OD
    Photometric accuracy (cuvette) < ±0.005 OD ±1.0%, 0–2 OD
    Photometric precision < ±0.003 OD ±1.0%, 0–2 OD
    Baseline flatness < 0.001 OD
    Stray light < 0.05% @ 230 nm
    Light source Xenon flash lamp
    Microplate read time (normal mode) 96 wells in 9 seconds 384 wells in 29 seconds
    Microplate read time (speed read mode) 96 wells in 5 seconds 384 wells in 16 seconds
    Temperature regulation
    Temp. range Ambient +4°C to 45°C
    Temp. uniformity (microplate) ±0.5°C at 37°C
    General specifications
    Dimensions (in.) 8.6 (H) x 22.8 (W) x 15 (D)
    Dimensions (cm) 22 (H) x 58 (W) x 38 (D)
    Weight 30 lbs. (13.6 kg)
    Power consumption < 250 watts
    Power source 100–240 Vac, 4 A 50/60 Hz
    Robot compatible Yes