Hach DREL/2400 portable touchscreen spectrophotometerHttps://www.hach.com/asset-get.download-en.jsa?id=7639984460
The only spectrophotometer designed for the water analysis industry that combines advanced optics, comprehensive testing and data management capabilities; simplified, touch-screen operation and complete portability. Users benefit from a high-performance optical system that boasts specifications typical of less-versatile laboratory units. A unique touch-screen user interface makes it easy to access more than 130 direct-reading tests and navigate through all analytical functions. Operators use battery or line power to put this powerful colorimetric analysis system to work anywhere throughout the process, in the field, or in the lab.
Automatic wavelength calibration and wavelength selection further simplify on-site analysis. The DR/2400 Portable Spectrophotometer identifies and logs up to 1000 data points for download via serial and parallel ports. It is available alone or as part of a complete portable laboratory including all apparatus and reagents for common water quality tests
With Hach's new DR/2400 Portable Spectrophotometer, you get:
The portable solution you need to optimize your water analysis program, anywhere, anytime.
The superior technology you expect in a spectrophotometer.
The ease of use you require to support all personnel.
The attractive price and value you want.
The DR/2400 Portable Spectrophotometer brings you the technically superior operation and innovative user interface of the popular DR/2500 Spectrophotometer, plus a design that makes it completely portable. Now you can have it all in your photometric water analysis program, whether it's municipal, industrial, environmental, or educational; in the lab, at the process site, or in the field.
Features and benefits
Automatic wavelength selection and calibration - Eliminate manual dialing and reduce errors.
"Favorite" programs menu -Store your frequently used tests to your own short list for easy access.
More than 130 preprogrammed tests - Just call up the test you need, each includes operator prompting selectable in four languages; select ""favorite"" programs quickly; add up to 50 single-wavelength, user-generated calibrations.
High-performance optical system - Achieve testing accuracy, instrument stability and longevity found in higher priced systems.
Graphic user interface with touch-screen display - Navigate easily through setup and testing with the touch of your finger; replaces the traditional keypad, repetitive scrolling, and keystroke combinations.
Comprehensive data management - Select time/date stamp, operator and sample ID; collect up to 1000 data points; download via RS-232 communication directly to printer or computer.
Battery-powered/compact size - Benefit from sophisticated spectrophotometer capability anywhere; line power converter included with instrument.
Programmed accuracy checks - Improve test technique and results with built-in, software-assisted standard additions checks.
Update capability - Keep your hardworking system updated with new Hach tests, using the plug-in Update Module programmed with new test calibrations; you add the calibration to your instrument memory without tedious keying sequences.
Accessories as you need them - Pay only for the options you need, including HachLinkâ„¢ Software for data formatting, the Pour-Thru Cell for convenient and accurate high-volume testing, and a variety of sample cells to match your requirements. Available alone or as a completely equipped portable laboratory: take your DR/2400 Portable Spectrophotometer wherever needed.
Specifications:
Wavelength Range 400 to 880 nm
Bandwidth 4 nm ± 1 nm
Wavelength Accuracy ± 1 nm
Wavelength Resolution 1 nm
Wavelength Selection Automatic, based on method selection
Optical System Concentric Spectrometer for multi-channel spectroscopy
Wavelength Calibration Automatic, via internal filter
Linear Photometric Range –2.000 to +2.000 Abs
Photometric Linearity ±0.005 Abs from 0.0 to 0.5 Abs; ±1% from 0.5 to 2.0 Abs.
Photometric Resolution 0.001 Abs, 0.1 %T
Recommended Recertification Interval One year (See 8.8 Recertification Interval .)
Stray Light >2.5 A, <0.3 %T @ 430 nm using 3 mm thickness GG-475 filter
Operational Modes Selectable; momentary (Display Lock on), or constant (Display Lock off)
Readout Modes Transmittance, absorbance, and concentration
RS232 Port Standard 9-pin bi-directional
Line Power 9 VDC @ 1 amp; 95 to 240 VAC; 50/60 Hz; automatic selection
Battery Power 3 “D” Cell Alkaline or Rechargeable Alkaline Batteries
Sample Compartment and
Cell Compatibility:
Note: For Hach Programs, be sure to use the cell type specified for program in the DR/2400 Portable Spectrophotometer Procedure ManualNo Cell Holder Required 1-in. (25.4-mm) round cells/vials, including AccuView
Round Cell Holder, 16-mm COD/Test ‘N Tube™ and UniCell™ vials Metric Rectangular Cell Holder (optional) 1-cm rectangular cells Multi-pathlength Cell Holder Pour-Thru Cell and multi-pathlength cells
Interface Graphic user interface 320 x 240 pixel graphic display with touchscreen
Data Storage 1000 data points (date, time, results, sample ID, user ID), 50 user calibrations
Dimensions 32.4 x 17.8 x 18.8 cm (12.75 x 7.0 x 7.4 in.)
Weight 2.1 kg (4.6 lbs)
Keypad 5 keys for power on/off, backlight on/off, contrast down, contrast up, and batter status
Printer External serial
Pour-Thru Cell 1-inch/1-cm multi-pathlength (optional)
Report Generation
Download stored information in a standard report format. Meets GLP guidelines, including a minimum of: date, time, sample identification tag, analyst initials (3), results, and instrument serial number.
Real-time Clock Yes
Real-time Calendar Yes
Operating Environment 0 to 40 °C (32 to 104 °F); 90% relative humidity, non-condensing
Storage Environment –20 to 60 °C (–4 to 140 °F); 85% relative humidity, non-condensing
Comes with: DR/2400 Spectrophotometer, Operations Manual and AC adapter